Filter | FITS Header Value | Pattern String | Focus Offset (µm) | Exptime for 20k counts (seconds) | Filter Central Wavelength (Å) | Filter Width (Å) |
---|---|---|---|---|---|---|
Non-Filter Positions | ||||||
OPEN | OPEN | OPEN | 0 | |||
DARK | DARK | DARK | N/A | |||
4 Hole Mask | 4HOLEMASK | 4Hole | ||||
Johnson-Cousins Filters | ||||||
U | U | U | +140 | 178 / 4* | ||
B | B | B | +105 | 9 | ||
V | V | V | +105 | 5 | ||
R | R | R | +140 | 3.5 | ||
I | I | I | +140 | 3.3 | ||
Sloan (SDSS) Filters | ||||||
SDSS u' | SDSS-U | SL-u | +180 | 313 / 4.5* | ||
SDSS g' | SDSS-G | SL-g | +180 | 4.5 | ||
SDSS r' | SDSS-R | SL-r | +180 | 3.5 | ||
SDSS i' | SDSS-I | SL-i | +180 | 4.25 | ||
SDSS z' | SDSS-Z | SL-z | +180 | 3.25 | ||
Other Broad Band Filters | ||||||
V+R | VR | VR | +185 | 6091.7 | 1763.9 | |
Yish | YISH | YISH | +105 | 26 | ||
General Narrow Band Filters | ||||||
[OIII] | OIII | OIII | +205 | 180 | 5001.7 | 25.5 |
Hα-On | HALPHAON | Ha-on | +155 | 115 | 6564.9 | 30.1 |
Hα-Off | HALPHAOFF | Ha-off | +155 | 37 | 6459.1 | 114.6 |
Wolf-Rayet Filters | ||||||
WC | WR-WC | WC | +165 | 115 | 4660.4 | 49.5 |
WN | WR-WN | WN | +160 | 4690.9 | 49.6 | |
CT | WR-CT | CT | +170 | 4756.1 | 54.6 | |
Comet Filters (Note: These filters are 4 inches round and suffer some vignetting on LMI) | ||||||
Ultraviolet Continuum | UC | UC | +175 | 500** / 80* | 3449 | 79 |
Blue Continuum | BC | BC | +175 | 130 / 3.5* | 4453 | 61 |
Green Continuum | GC | GC | +175 | 119 | 5259 | 56 |
Red Continuum | RC | RC | +175 | 70 | 7133 | 58 |
C2 | C2 | C2 | +175 | 46 | 5135 | 119 |
C3 | C3 | C3 | +175 | 4063 | 58 | |
CN | CN | CN | +175 | 480 / 14* | 3869 | 56 |
CO+ | CO+ | CO+ | +175 | 160 | 4266 | 64 |
H2O+ | H2O+ | H2O+ | +175 | 7028 | 164 | |
OH | OH | OH | +175 | 480** / 400* | 3097 | 58 |
NH | NH | NH | +175 | 3361 | 54 | |
Assorted Other Filters (inquire if you want to use, as they may not be readily available) | ||||||
W032 | W032 | W032 | +176 | 5300 | ||
5027 | 5027 | 5027 | +130 | 5027 | ||
K1593 | K1593 | K1593 | +194 | 5010 |
** OH & UC counts were at about the bias level for this exposure time using the CLSD lamps at 12V.
* Using the photo floods (a.k.a. IGRINS lamps) on the LMI lamp boom