  • Respond to Alarms (& Alerts)
    • Compressor calls First-Responder for help (email, txt)
      • Wake up
  • Determine Mimir Status and Electrical Power Status
    • Did compressor recover on its own? (~minutes)
    • How long has compressor been stopped?
      • Short time (< 1 hr)
      • Long time (> 1 hr)
    • Vacuum status?
      • Less than 10-4 torr = OK
      • Between 10-4 and 10-3
    • torrIf
      • torrIf compressor < 1 hr off, OK
        • If compressor > 1 hr off, Call Second-Responder
      • Higher than 10-3 torr = Call Second-Responder
    • Electrical Power Status – All 3-phases present?
  • Re-Evaluate and Intervene or Call Second-Responder
    • Get to Mesa
    • Re-Evaluate Mimir Vacuum Status & Electrical Power Status
      • If both OK, turn compressor on
      • If Vacuum NOT OK, call Second-Responder
      • If 3-Phase power not OK, wait…
  • Report Actions
    • Email to Dan Clemens about Status, Intervention
    • Txt to Dan Clemens if Second-Responder called
  • Problems beyond routine, CALL Dan Clemens (781-492-2726)