Mimir First-Responder Manual - 20160419 V1.1 DPC

Executive Summary

  • Respond to Alarms (& Alerts)
    • Compressor calls First-Responder for help (email, txt)
      • Wake up
    • Determine Mimir Status and Electrical Power Status
      • Did compressor recover on its own? (~minutes)
      • How long has compressor been stopped?
        • Short time (< 1 hr)
        • Long time (> 1 hr)
      • Vacuum status?
        • Less than 10-4 torr = OK
        • Between 10-4 and 10-3 torr
          • If compressor < 1 hr off, OK
          • If compressor > 1 hr off, Call Second-Responder
        • Higher than 10-3 torr = Call Second-Responder
      • Electrical Power Status – All 3-phases present?
    • Re-Evaluate and Intervene or Call Second-Responder
      • Get to Mesa
      • Re-Evaluate Mimir Vacuum Status & Electrical Power Status
        • If both OK, turn compressor on
        • If Vacuum NOT OK, call Second-Responder
        • If 3-Phase power not OK, wait…
    • Report Actions
      • Email to Dan Clemens about Status, Intervention
      • Txt to Dan Clemens if Second-Responder called
    • Problems beyond routine, CALL Dan Clemens (781-492-2726)

...

  • Temperature is 60 F
  • Input (return) gas pressure is 100 psi
  • Output (supply) gas pressure is 320 psi.

Compressor Status Web Page (Logger Webpage) under NORMAL operations

...